| Rotor, bucket, tube rack, table for the
model 2420 |
  |
| Rotor Name |
Nominal Capacity |
Tube Size (mm)
diam. x length |
Maximum Speed
(rpm) |
Maximum
RCF
(xg)(*1) |
Bucket Code No. |
Tube Rack Adaptor Code No. |
| Swinging Bucket Rotor |
RS-240
(*2) |
48 x
1.5/2mL(*3) |
9.5-11 x 36-60 |
4,000 |
1,950 |
053-5040 |
- |
| 40 x 6mL RIA
tube(*3) |
12.5-13.3 x 36-80 |
4,000 |
1,970 |
053-5030 |
- |
| 24 x 5-7mL narrow blood
collecting tube /10mL(*4,5) |
12.5-13.3 x 46-105 |
4,000 |
2,430 |
053-5850 |
- |
| 4 x 7-10mL blood collecting
tube/15mL(*6) |
15-17.2 x 86-110 |
4,000 |
2,540 |
053-4910 |
- |
| 4 x 15mL Conical
(TC) |
17 x 121 |
4,000 |
2,580 |
053-4910 |
- |
| 8 x 7-10mL blood collecting
tube/15mL(*6) |
15-17.2 x 86-110 |
4,000 |
2,520 |
053-4900 |
- |
| 8 x 15mL Conical
(TC) |
17 x 121 |
4,000 |
2,580 |
053-4900 |
- |
| 16 x 7-10mL blood collecting
tube/15mL(*6) |
15-17.2 x 86-110 |
4,000 |
2,500 |
053-4980 |
- |
| 24 x 7-10mL blood collecting
tube/15mL(*6) |
15-17.2 x 89-105 |
4,000(*7) |
2,490 |
053-7340 |
- |
| 4 x 7-10mL blood collecting
tube/15mL(*6) |
15-17.8 x 87-110 |
4,000 |
2,500 |
053-4990 |
055-7400(*8) |
| 4 x 50mL |
27-37 x 95-110 |
4,000 |
2,540 |
053-4990 |
- |
| 4 x 50mL Conical (TC)
|
30 x 117 |
4,000 |
2,560 |
053-5010 |
- |
| 4 x 15mL/50mL Conical
(TC) |
17 x 121/30 x 117 |
4,000 |
2,610/2,500
|
053-5010 |
055-6010(*9) |
TOP |
| Microplate Rotor |
| RMP-23 |
6 x Microplate |
86(W) x 128(D) x
52(H) |
4,000 |
1,790 |
Equipped with rotor |
- |
| 2 x Deep-well
Plate | | |
| (*1) |
To show the RCF exactly,
input the rotation radius. At the factory, the rotation
radius is pre-set at 14.6cm.
The above Maximum RCF is
the calculated value at the maximum speed at each
rotation radius. |
| (*2) |
Buckets and adaptors are
optional. |
| (*3) |
Tube support can be
produced for special tube sizes. Contact our
distributor. |
| (*4) |
Blood collecting tubes made by SEKISUI (Insepack
II), Termo (Venoject II), Nipro (Neotube) or Beckton
Dickinson (Vacutainer) can be used. |
| (*5) |
Upto 16tubes of 5/7mL Termo (Venoject II) tubes can
be used. |
| (*6) |
Blood collecting tubes made by SEKISUI (Insepack
II), Termo (Venoject II), Nipro (Neotube) or Beckton
Dickinson (Vacutainer) can be used. |
| (*7) |
In the case of 15mL taper tubes, please operate at
3,000rpm or less. Bucket 053-7340 is exclusively used
with 2420. |
| (*8) |
In case of using glass tube, rubber cushion 024-1400
is necessary. |
| (*9) |
This adaptor is used commonly for both 15mL and 50mL
conical tubes. |
TOP | | |