| Maximum
speed |
15,000rpm |
| Maximum
RCF |
21,040xg |
| Maximum
capacity |
1,020mL |
| Control
system |
Invertor
Microprocessor control, Speed, RCF, Time, Temperature,
Pre-cooling, FLASHING, Acceleration & Deceleration,
Memory (five channels) |
| Alarm
display |
Lid
open, Imbalance, Over speed, Rotor set state, abnormary
high temperature, Function for detecting an occurrence
of electrical abnormality in motor, inverter, speed
sensor and temperature sensor. |
| Acceleration/Deceleration |
Rapid,
Slow, Super slow |
| Speed
setting |
From
200 to 15,000rpm, 10rpm increments |
| Speed
indication |
Digital
display,
from 0 to 10rpm in 1rpm increments
from 10-15,500rpm in 10rpm increments |
| RCF
setting |
From
10 to 21,040xg, 10xg increments
|
| RCF
indication |
Digital
display, from 0 to 22,030xg in 10xg increments
|
| Integrator
setting and indication |
Digital
display, from 1.00 to 9.99 x109 g.sec |
| Timer
setting and indication |
| Digital
display, with Hold and Electric buzzer,
Memory flashing |
 |
from
1 sec. to 100 sec., in 1 sec. increments
setting and indication
from 100 sec. to 990 sec., in 10 sec.
increments setting and
1 sec. indication
from 1 min. to 100 min., in 1 min. increments
setting and indication
from 100 min. to 990 min., in 10 min.
increments setting and
1 sec. indication |
|
| Temperature
setting and indication |
| Digital
display, 1°C increments, Pre-cooling |
 |
1°C
increments setting from -20°C
to +40°C
1°C increments indication from
-20°C to +43°C |
| Control
the rotor temperature |
|
| Refrigerant |
R134a |
| Setting
the Sound that notify the end of the operation |
5 kinds
and Sound none |
| Power
requirements |
Single
phase AC 110/115V±10%, 50/60Hz, 15A
Single phase AC 220/230/240V±10%, 50/60Hz,
10A |
| Rated
voltage and Rated current |
110V,
9.1A / 115V, 8.7A
220V, 4.6A / 230V, 4.4A / 240V, 4.2A |
| Power
consumption and heat output |
0.78kW
2.8MJ/h |
| Dimensions
(WxDxH) and Weight |
700x650x360mm,
102kg |