| Rotor, bucket, tube rack, table for the
5500 model |
  |
| Rotor
Name |
Nominal
Capacity |
Tube
Size (mm)
diam. x length |
Maximum
Speed
(rpm) |
Maximum
RCF
(xg) |
Bucket
Code No. |
Tube
Rack Adaptor Code No. |
| Fixed
Angle Rotor |
| AT-2018M |
18
x 1.5/2mL |
9.5-11
x 36-42 |
15,000 |
20,630 |
- |
- |
| 18
x 0.5mL |
7.0-7.9
x 25-34 |
15,000 |
18,110 |
- |
055-7590 |
| 18
x 0.4mL |
5.4-6.1
x 26-48 |
15,000 |
20,120 |
- |
055-7580 |
| 18
x 0.2mL PCR |
5.9-6.3
x 20-24 |
15,000 |
15,850 |
- |
055-0500 |
| AT-2724M |
24
x 1.5/2mL |
9.5-11
x 36-42 |
15,000 |
20,630 |
- |
- |
| 12
x 0.5mL |
7.0-7.9
x 25-34 |
15,000 |
16,850 |
- |
- |
| 24
x 0.4mL |
5.4-6.1
x 26-48 |
15,000 |
20,380 |
- |
055-7580 |
| 18
x 0.2mL PCR |
5.9-6.3
x 20-24 |
15,000 |
14,840 |
- |
- |
AT-2730M
(Biohazard-proof) |
30
x 1.5/2mL |
9.5-11
x 36-42 |
14,000(*1) |
21,040 |
- |
- |
| 14
x 0.5mL |
7.0-7.9
x25-34 |
14,000(*1) |
17,530 |
- |
- |
| 30
x 0.4mL |
5.4-6.1
x26-48 |
14,000(*1) |
20,380 |
- |
055-1240 |
| 14
x0.2mL PCR |
5.9-6.3
x20-24 |
14,000(*1) |
16,430 |
- |
- |
TOP |
| Swinging
Bucket Rotor |
| ST-720M |
80
x 1.5/2mL |
9.5-11
x 36-42 |
3,500 |
2,290 |
053-5820 |
055-4800 |
| 48
x 5/7mL narrow blood collecting tube(*2) |
12-13.3
x 63-110 |
3,500 |
2,310 |
053-5820 |
055-4990 |
| 24
x 8mL CPT
cell preparation tube(*3) |
13.5-15.8
x 81-130 |
3,500 |
2,310 |
053-5820 |
055-6900 |
| 80
x 10mL |
12.5-13.8
x 65-95 |
3,500 |
2,260 |
053-5820 |
055-4740 |
| 48
x 5-10mL narrow or thick blood collecting tube/15mL
(*2) |
12-17.2
x 72-110 |
3,500 |
2,330 |
053-5820 |
055-0880 |
| 48
x 10m blood collecting tube/15mL |
15-17.2
x 81-110 |
3,500 |
2,330 |
053-5820 |
055-4750 |
| 68
x 5/7mL narrow blood collecting tube(*2) |
12.5-13.7
x 62-110 |
3,500 |
2,300 |
053-5840 |
055-0260 |
| 68
x 10mL blood collecting tube/15mL |
15-17.2
x 52-110 |
3,500 |
2,300 |
053-5840 |
- |
| 28
x 15mL Conical |
17
x 121 |
3,500 |
2,300 |
053-5840 |
- |
| 32
x 15mL Conical |
17
x 121 |
3,500 |
2,330 |
053-5820 |
055-4780 |
| 8
x 50mL |
27-36.4
x 85-110 |
3,500 |
2,330 |
053-5820 |
055-4760 |
| 16
x 50mL Conical |
30
x 117 |
3,500 |
2,330 |
053-5820 |
055-6070 |
| 4
x 100mL |
38-46.5
x 75-120 |
3,500 |
2,330 |
053-5820 |
055-4770 |
| 4
x 175mL
Conical(*4) |
61
x 118 |
3,500 |
2,330 |
053-5820 |
055-4850 |
| 4
x 200mL
Conical(*4) |
61
x 130 |
3,500 |
2,330 |
053-5820 |
055-4850 |
| 4
x 225mL
Conical(*4) |
61
x 137 |
3,500 |
2,330 |
053-5820 |
055-4850 |
| 4
x 250mL |
60-61.8
x 100-130 |
3,500 |
2,380 |
053-5820 |
055-4850 |
sputum
tube
(self-standing type)(*5) |
- |
- |
- |
- |
- |
TOP |
| Plate
Rotor |
| PT-21M |
6
x Micro-Plate |
86(W)
x 130(D) x 62(H) |
5,000 |
3,100 |
Equipped
with rotor |
- |
| 2
x Deepwell-plate |
| Medium
Capacity Fixed Angle Rotor |
| RA-410M3 |
8
x 12mL
Plastic tube(*12) |
15.9-16.5
x 97-120 |
10,000 |
12,970 |
- |
055-1120 |
| 8
x 14mL(*13) |
17
x 100 |
10,000 |
12,630 |
- |
055-0550 |
| 8
x 15mL Conical tube |
17
x 121 |
10,000 |
13,190 |
- |
055-1280 |
| 8
x 50mL
Plastic tube(*14) |
26.4-29.2
x 98-115 |
10,000 |
12,860 |
- |
055-1110 |
| 8
x 50mL Conical tube |
30
x 117 |
10,000 |
13,420 |
- |
- |
| Sartorius |
Vivaspin
2 |
16.7
x 129 |
10,000 |
13,190 |
- |
055-1280 |
| Vivaspin
6 |
16.7
x 129 |
10,000 |
13,190 |
- |
055-1280 |
| Vivaspin
20 |
30
x 117 |
10,000 |
13,420 |
- |
- |
| Millipore |
Amicon
Ultra-4 |
17
x 129 |
10,000 |
13,190 |
- |
055-1280 |
| Amicon
Ultra-15 |
30
x 121 |
10,000 |
13,420 |
- |
- |
TOP |
|
|
| (*1) |
Cooling down to 7°C
or below is possible at this speed. |
| (*2) |
5/7mL
narrow tube made by Sekisui (Insepack II), Termo (Venoject
II) Nipro (Neotube) and Becton Dickinson (Vacutenar) can
be used. Tube for flow cytometry can be used. |
| (*3) |
The CPT cell preparation
tube is made by Becton Dickinson (Vacutenar
8mL for CPT cell separation). |
| (*4) |
175 mL Falcon conical
tube (code no. 352076)
needs a cushion (code no. 352090).
200 mL Nalge Nunc conical tube (code no. 376813)
needs a cushion adapter (code no. 377585).
225 mL Falcon conical tube (code. no. 352075)
needs a cushion (code no. 352090). |
| (*5) |
Several types of sputum
tube will be acceptable with ST-720 and 053-5820 bucket.
For more information, please contact your local dealer. |
| (*7) |
The sealing adapter for
50mL conical (TC) (code no. 055-5760) , 15mL conical (TC)
(code no. 055-5750) and 50mL Oak ridge type tubes (code
no. 055-4310) can be used (Option). |
| (*8) |
15 mL glass tubes can
be used if an adapter (code no. 055-7400) is added and
need a rubber cushion (code no. 024-1400) sold separately. |
| (*9) |
15 mL conical (TC) tubes
can be used if an adapter (code no. 055-1280) is added. |
| (*10) |
If the microtiter plates
are used, it is possible to stack them in double or more
on each bucket. However, be sure to test the strength
of those plates prior to centrifugation. The microtiter
plate at the bottom may be crushed. |
| (*11) |
This adapter makes it
convenient to handle a micro titer plate , especially
some microtiter plates are stacked. |
| (*12) |
The NALGENE Round-Bottom
Centrifuge Tubes (Cat. No. 3110-0120
and 3117-0120
can be used. |
| (*13) |
The BD FalconTM
Round Bottom Test Tubes (Cat. No.352006,
352018
and 352059)
can be used. |
| (*14) |
The NALGENE Round-Bottom
Centrifuge Tubes (Cat. No. 3110-0120
and 3117-0120
can be used. |
TOP |
|
|