KUBOTAimage
Home
Product
+8920
  + Specification
  + Rotor List
  + Rotor, Bucket,
     Tube Rack, Table
  + More information
Company Overview
G-force Calculation
Contact Distributor
Contact Us
Site Map
ISO logo
KUBOTA's plant in Fujioka has become ISO9001 certified since 1996.
Rotor, bucket, tube rack, table for the 8920 model
Catalogue DownloadPDF logo

Rotor Name Maximum Capacity Tube Sizes (mm)
diam. x length
Maximum Speed
(rpm)
Maximum RCF
(xg)
Bucket Code No. Tube Rack
Adaptor Code No.
Swinging Bucket Rotor
RS-730
(*1)
80 x RIA tube 12 x 75-110 4,200 3,590 S13215 S18005
48 x 7mL blood collecting tube 12.5-13 x 75-106 4,200 3,590 S13215 S18012
48 x 15mL(*3) 15.5-17 x 96-115 4,200 3,590 S13215 S18015
16 x 15mL Conical (TC) 16.5 x 120.6 4,200 3,590 S13215 S18016
8 x 50mL Conical (TC)(*2) 29.5 x 117 4,200 3,590 S13215 S18052
160 x RIA tube(*4) 12 x 75-110 4,000 3,450 S13420 S18005
96 x 7mL blood collecting tube(*4) 12.5-13 x 75-106 4,000 3,450 S13420 S18012
96 x 15mL(*3,4) 15.5-17 x 96-115 4,000 3,450 S13420 S18015
40 x 15mL Conical (TC) 16.5 x 120.6 4,000 3,450 S13420 S18615
32 x 15mL Conical (TC)(*4) 16.5 x 120.6 4,000 3,450 S13420 S18016
24 x 50mL Conical (TC) 29.5 x 117 4,000 3,450 S13420 S18650
16 x 50mL Conical (TC)(*4) 29.5 x 117 4,000 3,450 S13420 S18052
28 x 50mL Conical (TC) 29.5 x 117 4,200 3,750 S13235 -
4 x 500mL(*6) 69 x 159 4,000 3,720 S13520 -
64 x RIA·7mL blood collecting tube 12-16 x 60-85 4,000 3,830 S13550 S17356
40 x 15mL attached caps 17-19.4 x 107 4,000 3,830 S13550 S17357
48 x 15mL 15.5-17 x 96-115 4,000 3,830 S13550 S17358
36 x 15mL Conical (TC) 16.5 x 120.6 4,000 3,830 S13550 S17359
12 x 50mL 34 x 100 4,000 3,830 S13550 S17375
16 x 50mL Conical (TC) 29.5 x 117 4,000 3,830 S13550 S17377
16 x 50mL Conical (TC) Self-Standing
29.5 x 117
4,000 3,830 S13550 S17378
4 x 250mL Bottle/230mL Glass tube(*7) 61.5 x 125/57 x 135 4,000 3,830 S13550 S17405/S17401
4 x 500mL 69 x 159 4,000 3,830 S13550 S13555
4 x 750mL 85 x 170 4,000 3,830 S13550 -
12 x Micro plate(*5) 86(W) x 130(D) x 49(H) 4,200 3,180 S13220 -
4 x Deep-well Plate(*6) 4,200 3,180 S13220 -
TOPTOP
Microtiter Plate Rotor
PT-20 6 x Micro plate(*5) 85(W) x 128(D) x 52(H) 4,000 1,790 Bucket included -
2 x Deep-well Plate(*5)
PT-21 6 x Micro plate(*5) 86(W) x 130(D) x 62(H) 5,000 3,100 Bucket included -
2 x Deep-well Plate(*5)
PT-89 8 x Micro plate(*5) 86(W) x 130(D) x 89(H) 2,400 1,080 Bucket included 055-6370
(*8)
4 x Deep-well Plate(*5)
PT-745 12 x Micro plate(*5) 86(W) x 128(D) x 90(H) 4,200 3,100 Bucket included 055-6380
(*9)
4 x Deep-well Plate(*5)
Fixed Angle Rotor
RA-288M3 24 x 12mL 15.9-16.5 x 97-120 8,500 9,690 - -
RA-410M3 8 x 15mL Conical (TC) 17 x 121 8,500 9,530 - 055-1280
8 x 50mL Conical (TC) 30 x 117 8,500 9,690 - -
RA-500M2 10 x 50mL 26.4-29.2 x 98-121 8,500 9,690 - -
(*1) Shielded covers are sold separately. The packet S13215 shield cover is divided into code No. S17112 (made of PC) and S17113 (made of PP). The packet S13420 shield cover is divided into code No. S17114 (made PC) and S17115 (made of PP). The packet S13550 sealing cap is code No. S17135 (made of PC).
(*2) FALCON®, SARSTEDT®, Lab-Tek®, GREINER (White), and IWAKI 50 mL conical tube can use shield covers.
(*3) When shielded covers are used, tubes cannot be inserted in the tube holes at the four corners of the bucket. Accordingly, the number of tubes for each bucket is reduced by four tubes. (Example: When 15 mL glass tubes are used in bucket S13420, up to 80 tubes can be used.)
(*4) The code No. S13420 bucket uses two tube racks each (a total of eight) in one bucket. Do not use this bucket when a different type of a tube rack is contained in this bucket to prevent a danger.
(*5) When micro plates are used, the plates exceeding two stages can be piled up in each bucket. However, use the plates after testing their strength before centrifugation. The micro plate at the lower stage may be crushed.
(*6) This bucket is divided into adapters (S13174 and S13175 combined jointly) for NALGEN® 175 mL type bottom bottle (3143-0175). Contact us.
(*7) 250 mL bottle adapter (S17405) and 230 mL glass tube adapter (S17401)
(*8) The adapters 055-6370 is available (Option). This adapter makes it convenient to handle a micro titer plate , especially some microtiter plates are stacked.
(*9) The adapters 055-6380 is requiered (Option).
TOPTOP

© Copyright 2002-2004, KUBOTA Corporation all rights reserved.